Don't have an account?
Dimensions (L × W × H)
20.866 in x 20.866 in x 54.015 in (530 mm x 530 mm x 1,372 mm)
Weight
Approx. 88 lbs (40 kg)
Power supply
Voltage rating 100-240 VAC / 50/60 Hz
Voltage supply (battery charging)
Voltage supply (battery charging)
Lithium-Ionen battery
running time ≥ 18 hours in operation
Media transport
Roll pump
Flow rate (with contrast media)
Can be set from 0.1 to 10.0 ml/s in increments of 0.1 ml/s (MRI)
Can be set from 0.1 to 4.0 ml/s in increments of 0.1 ml/s (contrast-enhanced mammography)
Injection protocol
Up to 40 Boli incl. pause
Maximum injection volume
1 to 10.0 ml, increment: 0.1 ml
10 to 100 ml, increment: 0.5 ml
100 to 400 ml, increment: 1 ml
Reservoir capacity
Contrast media: max. 2 × 200 ml (MRI), max. 2 × 250 ml
(contrast-enhanced mammography)
NaCl: max. 1 × 1,000 ml or max. 1 x 2,000 ml with a bag and optional media rod
Adjustable pressure limit
2-14 bar in increments of 0.1 bar (29-203 psi in increments of 1 ps1)
Air monitoring
3 detectors for the media supply tubing, 1 detector for the Easy-Click-Cassette and patient tubing
Pressure monitoring
1 pressure sensor
Maximum pressure limit
Peak-value: 203.1 PSI
tuneable from 29 – 203.1 PSI in 0,1 PSI steps
(Standard-pressure limit 159.5 PSI)
User interfaces
Terminal: 12” color TFT touch screen
Injector: 4” color TFT touch screen
Data import and export
WLAN (802.11 b/g/n)
Data transfer from injector to terminal
40 max.
Quantity of boli (phases) per injection program
200 max.
Number of injection programs that can be saved
200 max.
Pressure curve
Real-time display of the injection pressure curve
Software functions (standard)
CM-Loop
Display of remaining quantity
NaCl backup
Manual pause, time-controlled pause
Timer
Elapsed time
Software options
CM-Select
Injector accessories
Convenience package: Media rod, waste bin and tray
Disposables
Patient tubing 2.5 m or 3.2 m
Easy-Click-Cassette with holder s or holder l
Adapter ring for small bottles
MR Conditional: 0,55, 1,5 and 3 Tesla
Observe the safety distance to the gantry according to IFU